The instrument uses high precision constant current flowing through the measured parts and four end measurement, effectively eliminate the lead error; Using high precision
AD conversion, suitable for users to perform high precision measurement. The instrument has the sorting function (super, qualified, super) allow users to freely set the upper and lower limits of resistance and nominal values, greatly improve the test efficiency of the instrument
About VICTOR 6310:
Model number |
Test range |
Test range |
RS-232 interface |
HANDLER
|
VICTOR 6310 |
10uΩ-20KΩ |
200mΩ、2Ω、20Ω、200Ω、2KΩ、20KΩ |
Optional |
Optional |
﹝
2-2﹞Test range
VICTOR 6310: 10uΩ-20KΩ
﹝
2-3﹞Test range:
|
20mΩ |
200mΩ |
2Ω |
20Ω |
200Ω |
2kΩ |
20kΩ |
Accuracy |
0.2%±3 |
0.1%±3 |
|||||
Test
|
100mA |
100mA |
10mA |
1mA |
100uA |
100uA |
|
The reading
|
10uΩ |
100uΩ |
100uΩ |
mΩ |
10mΩ |
100mΩ |
1Ω |
﹝
2-4﹞显示The scope of:
Read straight: Take
The resistance value is disyed with 5 digits. If the first digit is greater than 2, only 4 digits are disyed
Percentage: To
4位数字加1符号位显示电阻百分比,显示The scope of为±9999%.
Overrange disy:
|_||_||_||_||_|
﹝
2-5﹞Test速度:
Quick: about
15 times per second
Slow speed: about
8 times per second
﹝
2-6] Trigger mode:
Single trigger
:
Instrumental
HANDLERInterface板(Optional)从外部接收到启动信号后,触发一次测量一次.也可通过面板上的触发键进行触发测量,每按键一次测量一次.若开机时为Single trigger则
5 digit digital tube disy -----.
连续触发:触发信号由仪器内部产生,并连续不断的对被Test件进行测量,并将结果输出显示到
5 digit digital tube
Three, parameter specifications
﹝
3-1] Rated working conditions:
(
1) Power supply voltage 220V±10percent frequency: 50Hz±5 percent 60 Hz
(
2) Working conditions: Temperature: 0℃ ~ 40℃, relative humidity: <80percentRH
(
3) Storage conditions: temperature: -10℃ ~ 70℃; Relative humidity: <70percentRH
(
4) Power: less than 30 W.
(
5) The phase line L, neutral line N and ground line E of the power supply should be the same as the power plug of the instrument, as shown in Figure 3.1.1. At the same time ground E should be reliably grounded, otherwise there may be leakage phenomenon on the surface of the instrument!
(
6). After the instrument is opened, it is necessary to preheat for 20 minutes, and then measure to ensure the accuracy of measurement
(
7) The power socket can not share a socket with high-power electrical equipment, so as not to interfere with the instrument work or electrical impact and damage the instrument