Product features
Ø Maximum measurement frequency of 100kHz,1Hz step continuously adjustable
Test level 10~2000mV,1mV step continuously adjustable
Support DC resistance (DCR), electrolytic capacitance measurement
Ø Internal bias voltage output 10mV~1500mV
Ø 3.5 "TFT disy, 5-bit disy
Ø Communication: USB Device,RS232 (or 485), Handle,GPIB(optional),USB Host(optional)
Data recording function (maximum, minimum, average value)
Ø Support for SCPI communication protocol
Provide system Settings, which can be configured as required in Chinese or English, buzzer, screen brightness, etc
Ø Basic measuring accuracy 0.2percent
Manual and automatic ranges
Ø With open circuit, short circuit calibration function
Ø Comparator sorting (5 gears), alarm function
Main technical indicators
Model number |
VICTOR 4090A |
VICTOR 4090B |
VICTOR 4090C |
VICTOR 4091A |
VICTOR 4091C |
Test frequency |
10 points (100120200400800, 1 K, 2 K, 4 K, 8 K, 10 K) |
12 points (100120200400800, 1 K, 2 K, 4 K, 8 K, 10 K, 15 K, 20 K) |
16 point (100120200400800, 1 K, 2 K, 4 K, 8 K, 10 K, 15 K, 20 K, 40 K, 50 K, 80 K, 100 K) |
10~10 KHZ continuous adjustable,1Hz step |
10~100 KHZ continuous adjustable,1Hz step |
The disy screen |
3.5寸TFT液晶The disy screen |
||||
Disy digit |
Primary parameter: 5-bit, auxiliary parameter: 5-bit |
||||
Measurement parameters |
Primary parameter: L/C/R/Z, auxiliary parameter: X/D/Q/θ/ESR |
||||
Measuring range (L) |
0.01μH - 9999H |
||||
Measuring range (C) |
0.01pF - 99.999mF |
||||
Measuring range (R) |
0.0001Ω - 99.999MΩ |
||||
Basic accuracy |
0.2% |
||||
Measure the speed of |
2 times per second (slow speed),4 times per second (medium speed),8 times per second (fast speed) |
||||
Internal bias |
0-1500mV adjustable,1mV step. |
||||
Test level |
Six fixed point level (0.1 V, 0.3 V, 0.6 V, 1 V, 1.5 V, 2 V) |
0.1V~2V continuously adjustable level,1mV step. |
|||
Source output impedance |
30Ω、100Ω |
||||
Calibration function |
Open circuit calibration, short circuit calibration |
||||
Filter function |
Filter limit range can be set - 50 percent ~ 50 percent, which fixed point percent for 1, 5 percent, percent 10, 20 percent |
||||
Comparator sorting |
Fifth grade sorting ,三档合格,一档不合格,一个附属档 |
||||
Communication interface |
Standard configuration RS232 (or 485),USB Device,Handle, optional GPIB,USB Host |
||||
Other |
Support DC resistance (--) DCR), electrolytic capacitance measurement mode, backlight brightness adjustable, Chinese, English optional |
General technical specification
Power supply voltage: 220V.AC±10percent,110V.AC±10percent (optional),45~65Hz
Ø Power consumption: < 20W
Ø Disy: 3.5 "TFT LCD screen, resolution 480×320, color 16M
Temperature range: 10℃~ 40℃ in operating state, -10℃~ 60℃-- in non-operating state
Ø Humidity range: 0~40℃,≤ 90% relative humidity
Ø 接口:Standard configurationRS232 (or 485),USB Device,Handle, optional GPIB,USB Host