Product specifications: Fluke 369/CN and 369 FC/CN microampere true RMS leakage current clamp meter
Electrical technical specifications | ||
Measuring function | Ac current | |
Ac current量程 | 3 mA、30 mA、300 mA、3 A、30 A、60 A | |
Range selection | mA/A: Manually select
3 mA / 30 mA / 300 mA: Automatically selects 3 A / 30 A / 60 A: Automatically selects |
|
Range/resolution | 3 mA¹ / 0.001 mA | |
30 mA / 0.01 mA | ||
300 mA / 0.1 mA | ||
3 A / 0.001 A | ||
30 A / 0.01 A | ||
60 A / 0.1 A | ||
Accuracy -- Activate filter (40 to 70 Hz), shut filter (40 to 1 kHz) ² | 3 mA – 30 A
60 A |
1percent 5 digits
2percent 5 digits |
Frequency | 40hz - 1 kHz | |
Peak coefficient | ="3" | |
After calibration, the accuracy is valid for one year. Accuracy is expressed as ± (percent number of readings). The reference temperature condition is 23 ± 5 °C and the maximum relative humidity is 80percent
¹ The minimum value is 10 μA true effective value 2 . |
||
The characteristics of the body | ||
Disy screen (LCD) | Digital reading: 3300 words | |
Disy update rate | 4 times per second | |
Maximum measured conductor diameter | 61 mm | |
Size | 257 x 116 x 46 mm | |
The weight of | 600 g | |
Battery | 2 AA,IEC LR6,NEDA 15A, Basic | |
Battery寿命 | 在不使用背光灯和聚光灯的情况下,预计Battery寿命可以过 150 小时 | |
Automatic shutdown | 钳表将在 15 分钟无操作后Automatic shutdown | |
Safety and environmental norms | ||
General safety | IEC 61010-1: Pollution Level 2 | |
Measurement safety | IEC 61010-2-032:CAT III 600 V / CAT IV 300 V | |
Operating temperature | -10°c - 50°c | |
Storage temperature | -40°c - 60°c | |
Operating humidity | No condensation (< 10 °C) | |
90percent relative humidity (10 °C to 30 °C) | ||
75percent relative humidity (30 °C to 40 °C) | ||
45percent relative humidity (40 °C to 50 °C) | ||
Protection level | IEC 60529: IP30 (Jaws closed) | |
Working altitude | 2000 m | |
Storage elevation | 12000 m | |
Current sensor action class | IEC 61557-13: 1 μ g,≤100a / m | |
Electromagnetic compatibility (EMC) | ||
International standard | ||
IEC 61326-1 | Industrial electromagnetic environment | |
CISPR 11 | Group 1 Category B | |
The first group | The device needs to generate and/or use radiofrequency energy associated with conducted energy within the device, which is essential for its own internal function | |
Grade B | The device is suitable for household appliances as well as any device directly connected to a low voltage network powering a residential building. When connected to the test target, the device may exceed the launch stage specified in CISPR 11 requirements
Korea (KCC) : Category A Equipment (Industrial Broadcasting and communications equipment) |
|
Grade A | This product meets the requirements of industrial electromagnetic equipment, sellers or users should pay attention to. This device is designed for use in a commercial environment, not a residential one
USA (FCC) : 47 CFR 15 Part B. This product is deemed to be duty-free equipment in accordance with Clause 15.103 |