  • Fluke 1625-2 grounding resistance tester

    Model:Fluke 1625-2

    Brand:FLUKE

    Introduction:Fluke 1625-2 is a distinctive grounding tester that can complete all four types of grounding measurement: Tripole and quadrupole potential drop (using ground pile) Quadrupole soil resistivity test (using ground pile) Selective test (using ground pile and 1 clamp) Pileless test (using 2 clamp only)

  • Product specifications: Fluke 1625-2 grounding resistance tester

    General technical index
    Memory 通过 USB 端口可访问内部Memory存储的多达 1500 条记录
    Measuring function Interference voltage and frequency, tripole and quadrupole ground resistors with/without clip current transformers, dipole AC resistors, dipole and quadrupole DC resistors
    The disy screen 4 数位(2999 数位) - 7 段液晶The disy screen(可视性提高)
    Operation Central rotary switch and function keys
    Temperature range
    Operating temperature -10°c - 50°c(14°f - 122°f
    Storage temperature -30°c - 60°c(-22°f - 140°f)
    Temperature coefficient ±0.1% /°C <18°C >28°C——
    Protection type According to EN60529, the shell protection class is IP56; The protection grade of the battery cover is IP40
    Maximum voltage Warning - Jack" Jaws "; To the socket E,ES,S or H
    Urms = 0 V
    插口 E、ES、S 或 H 相互之间任意组合,Urms = 250 V (misuse)
    Security Double and/or enhanced insulation protection. Maximum ground to 50 V (according to IEC61010-1). Pollution Level 2
    Quality standard Developed, designed and manufactured in accordance with DIN ISO 9001
    Outfield influence In accordance with DIN 43780 (8/76)
    Auxiliary power supply 6 x 1.5V alkaline battery (IEC LR6 or AA type)
    Battery life range 使用 IEC LR6/ AA 型:通常为3,000 次测量 (RE +RH ≤ 1 kΩ)
    使用 IEC LR6/ AA 型:通常为6,000 次测量 (RE + RH > 10 kΩ)
    Dimensions (width x height x depth) 250 x 133 x 187 mm (9.75 x 5.25 x 7.35 in)
    The weight of ≤ 1.1 kg (2.43 lb) (without accessories); 7.6kg (16.8lb) (accessories and battery included in carrying case)
    Shell material Polyester
    干扰电压直流 + 交流的测量 (UST )
    Measurement error limits: Methods Full wave rectifier
    Measuring range 1 v - 50v
    Disy range 0.0 v - 50v
    Resolution 0.1 V
    Frequency range Ac/DC sine wave 45 Hz to 400 Hz
    Precision ± (5percent of the reading in 5 digits)
    Measurement sequence About 4 measurements per second
    Internal resistance * * 1.5 mΩ——
    Maximum overload Urms = 250 V
    Measurement of interference frequency (F)
    Measurement error limits: Methods Measure the oscillation period of the interference voltage
    Measuring range 6.0 Hz- 400 Hz
    Disy range 16.0 Hz、299.9 Hz、999 Hz——
    Resolution 0.1 Hz- 1 Hz
    The range of 1 v - 50v
    Precision ± (1percent of the reading in 2 digits)
    Ground resistance (RE)
    Measuring method Measuring current and voltage with a probe according to IEC61557-5
    Open-circuit voltage 20/48 V, ac
    Short circuit current 250 mA (AC)
    Measure the frequency Manually or automatically select 94,105,111 and 128 Hz(AFC), which in the function R¹ is 55 Hz
    Noise suppression 120 dB(16 2/3、50、60、400 Hz)
    对地Maximum overload Urms = 250 V
    Electrical measurement specifications
    The amount of inherent error or influence Reference conditions or specified work scope Symbol code In accordance with the IEC1557 standard relevant part of the requirements or tests Test type
    Inherent error Reference conditions A Section 6.1 of Part 5 R
    Position 参考Position ±90 ° E1 Part 1, Section 4.2 R
    Supply voltage In accordance with the manufacturer's limits E2 Part 1 Sections 4.2 and 4.3 R
    Temperature 0°c ~ 35°c E3 Part 1, Section 4.2 T
    Series interference voltage See sections 4.2 and 4.3 E4 Part 5 Sections 4.2 and 4.3 T
    Resistance of the probe and auxiliary grounding electrode 0 至 100 x RA But ≤ 50 kΩ E5 Part 5, Section 4.3 T
    System frequency 99percent to 101percent of the nominal frequency E7 Part 5, Section 4.3 T
    System voltage Nominal voltage 85percent to 110percent E8 Part 5, Section 4.3 T
    Operation误差 B = ±(|A| + 1,15 √E²123456 ) Part 5, Section 4.3 R
    B[percent] = ± B/ reference value x 100percent

    A = Inherent error
    En = Deviation
    R = Routine test
    T = type test
    Measuring range 0.020 Ω清洁度300 kΩ——
    Disy range 0.001 Ω清洁度2.999 Ω——
    3.00 Ω 29.99 Ω——
    30.0 Ω清洁度299.9 Ω——
    0.300 kΩ清洁度2.999 kΩ——
    3.00 kΩ 29.99 kΩ——
    30.0 kΩ清洁度299.9 kΩ——
    Resolution 0.001 Ω
    0.01 Ω
    0.1 Ω
    1 Ω
    10 Ω
    100 Ω
    Precision ± (2percent of the reading in 2 digits)
    Operation误差 ± (5percent of the reading in 5 digits)
    Measure the time 固定频率时通常为 8 秒使用自动频率控制, 30 秒,完成有Measure the frequency循环
    由于Resistance of the probe and auxiliary grounding electrode导致的附加误差 RH (RS + 2000 Ω)/RE x 1.25 x 10-6 percent five digits
    RH 和 RS Measurement error 固定频率时通常为RE + RS + RH 10percent of
    Maximum probe resistance ≤ 1 MΩ
    Maximum auxiliary grounding electrode resistance ≤ 1 MΩ

    If the error is within the limits of IEC61557-5, it will be checked automatically
    如果由于某些条件的影响,导致测量探针、辅助接地电极和接地电阻后的Measurement error过 30%,The disy screen将显示警告符号并提示 RS 或 RH 过高.

    测量Resolution的自动转换取决于辅助接地电极电阻 RH
    Umeas RH-- at 48 V <300 Ω
    <6 Ω
    <60 Ω
    <600 Ω
    Umeas Is equal to 20 V RH <250 Ω
    <2.5 kΩ
    <25 kΩ
    <250 kΩ
    Resolution 1 mΩ
    10 mΩ
    100 mΩ
    1 Ω
    选择性接地电阻测量 (RE The mouth of the pliers)
    Measuring method The current and voltage are measured with a probe according to EN61557-5 and the current in a single branch is measured with an additional transformer
    Open-circuit voltage 20/48 V, ac
    Short circuit current 250 mA (AC)
    Measure the frequency Manual or automatic options 94,105,111 and 128 Hz (AFC),55 Hz (R¹)
    Noise suppression 120 dB(162/3、50、60、400 Hz)
    对地Maximum overload 对地Urms = 250 V (will not start measurement)
    Measuring range 0.020 Ω清洁度300 kΩ——
    Disy range 0.001 Ω清洁度2.999 Ω——
    3.00 Ω 29.99 Ω——
    30.0 Ω清洁度299.9 Ω——
    0.300 kΩ清洁度2.999 kΩ——
    3.00 kΩ 29.99 kΩ——
    Resolution 0.001 Ω
    0.01 Ω
    0.1 Ω
    1 Ω
    10 Ω
    Precision ± (7percent of the reading in 2 digits)
    Operation误差 ±(读数的 10percent five digits)
    Additional errors due to typical electrode resistance of the probe and auxiliary grounding RH (RS + 2000 Ω)/RETOTAL x 1.25 x 10-6 percent five digits
    RH 和 RS Measurement error 固定频率时通常为RETOTAL + RS + RH 10percent of
    Measure the time 固定频率时通常为8 秒使用自动频率控制, 30 秒,完成有Measure the frequency循环.
    Minimum current of a single branch to be measured 0.5 mA Use the transformer (1000:1)
    0.1 mA Use the transformer (200:1)
    Maximum interference current through the transformer 3 A Use the transformer (1000:1)
    1. Use the recommended current pliers/transformer.
    Resistance measurement (R~)
    Measuring method Measure current and voltage
    Measure voltage 20 V AC voltage, rectangular pulse
    Short circuit current > 250 mA (AC)
    Measure the frequency Manual or automatic options 94,105,111 and 128 Hz (AFC)
    Measuring range 0.020 Ω清洁度300 kΩ——
    Disy range 0.001 Ω清洁度2.999 Ω——
    3.00 Ω 29.99 Ω——
    30.0 Ω清洁度299.9 Ω——
    300 Ω清洁度2999 Ω——
    3.00 kΩ 29.99 kΩ——
    30.0 kΩ清洁度299.9 kΩ——
    Resolution 0.001 Ω
    0.01 Ω
    0.1 Ω
    1 Ω
    10 Ω
    100 Ω
    Precision ± (2percent of the reading in 2 digits)

    Operation误差 ± (5percent of the reading in 5 digits)
    Measure the time Usually 6 seconds
    Maximum interference voltage 24 V, more than 24 V can not start measurement
    Maximum overload Urms The maximum value is 250 V
    Resistance measurement (R DC)
    Measuring method According to IEC61557-4 current - voltage measurement
    Measure voltage Dc 20 V
    Short circuit current 250 mA DC
    The formula for measuring the value For quadrupole measurement, the H,S,ES wire can be extended without additional error
    Resistance > in wire E; At 1 Ω, an additional error of 5m Ω/Ω can be caused
    Measuring range 0.020 Ω清洁度300 kΩ——
    Disy range 0.001 Ω清洁度2.999 Ω——
    3.00 Ω 29.99 Ω——
    30.0 Ω清洁度299.9 Ω——
    300 Ω清洁度2999 Ω——
    3.0 kΩ清洁度29.99 kΩ——
    30.0 kΩ清洁度299.9 kΩ——
    Resolution 0.001 Ω
    0.01 Ω
    0.1 Ω
    1 Ω
    10 Ω
    100 Ω
    Precision ± (2percent of the reading in 2 digits)
    Operation误差 ± (5percent of the reading in 5 digits)
    Measurement sequence About 2 measurements per second
    Measure the time Usually 4 seconds, including polarity reversal (bipolar or quadrupole)
    Maximum interference voltage Dc or AC voltage ≤3 V, more than 3 V can not start measurement
    Maximum induction rate 2 Henry
    Maximum overload Urms = 250 V
    测量导线电阻的补偿 (RK )
    在函数 RE 三极、RE 四极(钳口)、R 交流和 R 直流二极中,可以打开测量导线电阻的补偿 (RK )
    The formula for measuring the value R显示 = R测量 - R补偿 ²
    2. Set point input value RK = 0.000 Ω, by adjusting the measurement in the range of 0.000 to 29.99 Ω
    Pileless grounding circuit measurement (double clamp, pileless)
    Switch gear RA quadrupole (double clamp, no pile)
    Resolution 0.001 Ω清洁度0.1 Ω——
    Measuring range 0.02 Ω 199.9 Ω——
    Precision ± (7percent of 3 digits of the reading)
    Operation误差 ±(读数的 10percent five digits)
    Measure voltage Vm = 48 V AC (primary voltage)
    Measure the frequency 128 Hz
    Noise current (IEXT) Max IEXT = 10 A (AC) (RA < 20 Ω)
    Maximum to ground IEXT = 2 A (AC) (RA > 20 Ω)
    Measurement principle: Two current transformers are used to measure the resistance in the closed loop without piles
    自动The range of选择.
    Information about a dockless ground loop measurement is valid only if the measurement is made using the recommended current clamp at the specified minimum distance