testo 327-1烟气分析仪,包括充电电池和出厂报告,测量O2, CO, hPa和℃.
Product Introduction:
testo 327-1是您的烟气分析入门仪器.这台仪器测量℃,O2 ,CO2 ,CO and flue gas extraction and other combustion efficiency parameters. Powerful LED backlit disy and four lines of parameter disy ensure that you can easily read in harsh environments. The device stands out for its easy-to-understand menu, streamlined case design and durability
Advantages at a glance:
The menu is simple and easy to understand
Four parameters are disyed at the same time
LED backlight disy
Fast infrared printing
Built-in condensing tank
TÜV By RgG 253 acc. to
1. BImSchV
O2 ,℃,hPa in line with EN 50379 Part 2
CO complies with EN 50379 Part 3
Lightweight rechargeable lithium battery (1200mA, 5 hours), can be charged in the instrument or external charger
Integrated probe interface, convenient connection probe
The flue gas probe can measure the ambient CO concentration
管道夹层O2 Concentration measurement (can be stored)
Independent ambient temperature measurement
Undiluted CO measurement (can be stored)
Suction force measurement
6-8 fuels (depending on the national version, such as 6 in the UK version)
IP 40
Technical parameters |
|
---|---|
Storage temperature |
-20 ~ +50 °C |
Operating temperature |
-5 ~ +45 °C |
Battery life |
> 5 h |
The weight of |
About 500 g |
Size |
216 x 68 x 47 mm |
Type K thermocouple |
|
---|---|
The range of |
-40 ~ +600 °C |
Precision |
±0.5 °C (0.0 ~ +99.9 °C)
±0.5 percent of the measured value (100 ~ 600 °C) |
Resolution |
0.1 °C |
Suction force measurement |
|
---|---|
The range of |
±40 hPa |
Precision |
±0.02 hPa (-0.50 ~ +0.60 hPa)
±0.03 hPa (+0.61 ~ +3.00 hPa) ±1.5percent (3.01 ~ 40 hPa) |
Smoke loss |
|
---|---|
The range of |
0 ~ 99.9% |
Resolution |
0.1% |
Electrochemical CO |
|
---|---|
The range of |
0 ~ 4000 ppm |
Precision |
±20 ppm (0 ~ 400 ppm)
±5percent (401 ~ 1000 ppm) ±10percent of the measured value (1001 ~ 4000 ppm) |
Resolution |
1 ppm |
电化学O2 | |
---|---|
The range of |
0 ~ 21 Vol. % |
Precision |
±0.2 Vol. % |
Resolution |
0.1 Vol. % |
Combustion efficiency |
|
---|---|
The range of |
0 ~ 120% |
Resolution |
0.1% |
CO2 Calculate |
|
---|---|
The range of |
0 ~ CO2 max |
Precision |
±0.2 Vol. % |
Response time t90 |
About 40 s |